posted on 2015-05-14, 13:29authored byB. S. Bhatia, S. L. Bugby, J. E. Lees, A. C. Perkins
Existing protocols for assessing the performance characteristics of large field-of-view (LFOV) gamma cameras can be inappropriate and require modification for use with small field-of-view (SFOV) gamma camera systems. This communication proposes a generic scheme suitable for evaluating the performance characteristics of SFOV gamma cameras, based on modifications to the standard procedures of NEMA NU1-2007. Key differences in methodology between tests for LFOV and SFOV gamma cameras are highlighted along with the rationale for these changes. It is envisaged that this scheme will provide more appropriate methods for equipment characterisation, ensuring quality and consistency for all SFOV cameras.
Funding
This work is supported by an STFC CLASP Grant (ST/I00327/4).
SLB is in receipt of an STFC studentship
History
Citation
Physica Medica, 2015, 31 (1), pp. 98-103
Author affiliation
/Organisation/COLLEGE OF SCIENCE AND ENGINEERING/Department of Physics and Astronomy
Version
VoR (Version of Record)
Published in
Physica Medica
Publisher
Elsevier for Associazione Italiana di Fisica Medica, European Federation of Organisations for Medical Physics