posted on 2014-03-12, 13:48authored bySarah L. Bugby, John E. Lees, B. S. Bhatia, A. C. Perkins
A handheld, high-resolution small field of view (SFOV) pinhole gamma camera has been characterised using a new set of protocols adapted from standards previously developed for large field of view (LFOV) systems. Parameters investigated include intrinsic and extrinsic spatial resolution, spatial linearity, uniformity, sensitivity, count rate capability and energy resolution. Camera characteristics are compared to some clinical LFOV gamma cameras and also to other SFOV cameras in development.
Funding
Science and Technologies Facilities Council studentship and an STFC CLASP grant (ST/I00327/4).;Open Access funded by Science and Technology Facilities Council
History
Citation
Physica Medica, 30 (2014) 331-339
Author affiliation
/Organisation/COLLEGE OF SCIENCE AND ENGINEERING/Department of Physics and Astronomy
Published in
Physica Medica
Publisher
Elsevier on behalf of Associazione Italiana di Fisica Medica